Model Number 44010 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/23/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient weight not available from the site.No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a ventricular port.After implanting this liquid storage bag into the patient, the tube leaked liquid.The doctor asked the patient to replace the liquid storage bag.The procedure was completed with backup products.
|
|
Manufacturer Narrative
|
H2) additional information was added to the event description.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information: this issue was involving a liquid storage sac, and there was no such thing as a liquid storage bag.The sales article should not have mentioned it.The doctor asked the patient to replace the liquid storage sac.
|
|
Manufacturer Narrative
|
H3) the catheter associated with the port system was returned.There was a tear on the catheter, not clearly visible under the "vis." however, during leakage tests, air bubbles were noticed near the top half of the catheter and hence the catheter did not meet requirements for leakage.Instructions for use (ifu) cautions, ¿low tear strength is a characteristic of most silicone elastomer materials.Improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting or crushing of components." medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|