• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH BATTERY REAMER/DRILL FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH BATTERY REAMER/DRILL FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.705
Device Problems Separation Failure (2547); Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2022
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
 
Event Description
It was reported from japan that the battery reamer device was unable to connect to the battery firmly and the battery came off without pressing the button.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown but was noted to have occurred in 2022.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During evaluation it was determined that no issues with the device were identified.All pre-repair diagnostic assessment tests were passed.Therefore, the reported condition could not be duplicated nor confirmed.An assignable root cause was not determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY REAMER/DRILL FOR BPL II
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key14498305
MDR Text Key300445791
Report Number8030965-2022-03486
Device Sequence Number1
Product Code GEY
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/25/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.705
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received06/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/10/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-