Model Number N/A |
Device Problems
Misassembled (1398); Component Missing (2306); Structural Problem (2506); Calibration Problem (2890); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/21/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Telephone number: (b)(6).Evaluation and investigation is in process.Once the investigation is complete, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the dermatome malfunctioned outside of the surgical environment.Additional information received is that it was during maintenance the following failure descriptions were noted: control bar issue; calibration issue; damaged machined head need to be replaced; missing screw need to be replaced.The dermatome had a screw missing, a reportable malfunction.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This follow-up is being submitted to relay additional information.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Event Description
|
The customer has sent both articles back, to see if it is the combination or connection problem.No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up is being submitted to relay additional information.D10 - medical product: zimmer air dermatome ii hose: synthes (ao), 3 m catalog #: 00885100203 lot #: unk serial#: (b)(6).The results of the investigation are as follows: review of the most recent repair record determined the machined head was damaged, a screw was missing, the unit was out of calibration and the position of the control bar was not correct.The machined head, screw, shaft and sleeve bearings were replaced and the control bar was recalibrated and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|