Model Number N/A
Device Problems Misassembled (1398); Component Missing (2306); Structural Problem (2506); Calibration Problem (2890); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/21/2022
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Telephone number: (b)(6).Evaluation and investigation is in process.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that the dermatome malfunctioned outside of the surgical environment.Additional information received is that it was during maintenance the following failure descriptions were noted: control bar issue; calibration issue; damaged machined head need to be replaced; missing screw need to be replaced.The dermatome had a screw missing, a reportable malfunction.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
This follow-up is being submitted to relay additional information.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Event Description
The customer has sent both articles back, to see if it is the combination or connection problem.No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up is being submitted to relay additional information.D10 - medical product: zimmer air dermatome ii hose: synthes (ao), 3 m catalog #: 00885100203 lot #: unk serial#: (b)(6).The results of the investigation are as follows: review of the most recent repair record determined the machined head was damaged, a screw was missing, the unit was out of calibration and the position of the control bar was not correct.The machined head, screw, shaft and sleeve bearings were replaced and the control bar was recalibrated and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DERMATOME AN HANDPIECE
Type of Device
DERMATOME HANDPIECE, PNEUMATIC
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key14499462
MDR Text Key292725410
Report Number0001526350-2022-00502
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024464834
UDI-Public(01)00889024464834(11)190204(10)641310247
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/25/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number88710100
Device Lot Number641310247
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2022
Was the Report Sent to FDA? No
Date Manufacturer Received09/14/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/04/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-