MEDTRONIC, INC. PERCEPTA CRT-P MRI SURESCAN; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number W1TR01 |
Device Problems
Electrical /Electronic Property Problem (1198); Capturing Problem (2891)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 05/06/2022 |
Event Type
Injury
|
Event Description
|
It was reported that the patient was experiencing extracardiac stimulation from the cardiac resynchronization therapy pacemaker (crt-p) that was reproducible on the atrial lead.The device feature that automatically monitors and adjusts atrial pacing thresholds at periodic intervals was turned off, but the patient continued to have sensations of stimulation in the middle of the night.It was suspected that the stimulation was also related to the atrial lead position check (alpc).The crt-p remains in use.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|