Model Number N/A |
Device Problem
Failure to Cut (2587)
|
Patient Problems
Abrasion (1689); Failure of Implant (1924); Insufficient Information (4580)
|
Event Date 04/25/2022 |
Event Type
Injury
|
Event Description
|
It was reported that the skin was taken irregularly, patient skin damaged.The event timing was during surgery.
|
|
Manufacturer Narrative
|
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.
|
|
Manufacturer Narrative
|
Review of the most recent repair record determined that the reported issue could not be confirmed, however, the motor was not running and the unit was out of calibration and the zero reading.The motor, plug harness assembly, eccentric shaft, switch mount and switch were replaced to resolve issues related to the complaint.-a definitive root cause cannot be determined.-the event cannot be confirmed if any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trend.
|
|
Event Description
|
There is no information regarding whether an additional graft was required.No additional information is available.
|
|
Event Description
|
No additional event information available.
|
|
Manufacturer Narrative
|
If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|