• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME; DERMATOME, LINE-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME; DERMATOME, LINE-POWERED Back to Search Results
Model Number N/A
Device Problem Failure to Cut (2587)
Patient Problems Abrasion (1689); Failure of Implant (1924); Insufficient Information (4580)
Event Date 04/25/2022
Event Type  Injury  
Event Description
It was reported that the skin was taken irregularly, patient skin damaged.The event timing was during surgery.
 
Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.
 
Manufacturer Narrative
Review of the most recent repair record determined that the reported issue could not be confirmed, however, the motor was not running and the unit was out of calibration and the zero reading.The motor, plug harness assembly, eccentric shaft, switch mount and switch were replaced to resolve issues related to the complaint.-a definitive root cause cannot be determined.-the event cannot be confirmed if any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trend.
 
Event Description
There is no information regarding whether an additional graft was required.No additional information is available.
 
Event Description
No additional event information available.
 
Manufacturer Narrative
If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME, LINE-POWERED
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key14517947
MDR Text Key292792220
Report Number0001526350-2022-00511
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 09/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number63138521
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Date Manufacturer Received09/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/27/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-