Model Number N/A
Device Problems Inability to Irrigate (1337); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/02/2022
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Telephone number: (b)(6).Investigation is complete.This is an initial final report submission.Functional testing of the returned product identified the suction was not working and the motor made a high pitched noise.Visual examination of the interior of the hand piece found the motor mounts were melted and warped, dropping the motor out of position and preventing the pinion gear from making contact with the face gear.There were brown burn marks found on the corners of the motor mounts.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that during surgery the product pump was not working and would not spray and an unusual noise occurred when switched on.No harm or delay reported.At product evaluation brown burn marks found on the corners of the motor mounts.No adverse events were reported as a result of this malfunction.No additional event information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAN SPRAY KIT
Type of Device
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key14522847
MDR Text Key292833084
Report Number0001526350-2022-00519
Device Sequence Number1
Product Code FQH
UDI-Device Identifier00889024375154
UDI-Public(01)00889024375154(17)240324(10)65058749
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial
Report Date 05/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/24/2024
Device Model NumberN/A
Device Catalogue Number00515047500
Device Lot Number65058749
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/17/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/25/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-