MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS FEM COMPONENT, CEMENTED; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS FEM COMPONENT, CEMENTED; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5515-F-501
Device Problems Loss of Osseointegration (2408); Malposition of Device (2616)
Patient Problems Ambulation Difficulties (2544); Inadequate Osseointegration (2646)
Event Date 05/10/2022
Event Type  Injury  
Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.
 
Event Description
Revision of ps femur to ts femur with augments.Femoral component had changed alignment since original surgery in (b)(6) 2021 and is now in 10deg valgus according to precision navigation.On x-ray the surgeon stated the alignment now look valgus.Tibial components not revised.Ts liner implanted.Left side.
 
Event Description
Revision of ps femur to ts femur with augments.Femoral component had changed alignment since original surgery in (b)(6) 2021 and is now in 10deg valgus according to precision navigation.On x-ray the surgeon stated the alignment now look valgus.Tibial components not revised.Ts liner implanted.Left side.
 
Manufacturer Narrative
Reported event: an event regarding loosening involving a triathlon femoral component was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: a review of the provided medical records by a clinical consultant indicated: this inquiry concerns a patient who had a primary cemented triathlon ps total knee arthroplasty with a stem extension on the tibia.Less than one year later the patient required revision for a "change in position" of the femoral component.The surgeon elected to use a non-stryker cement.I can only assume that the "change in position" of the femoral component was due to early loosening.There was nothing in the history to suggest trauma.In my experience, early loosening of a cemented femoral component is uncommon.It can occur if infection is present but aseptic loosening is unusual.I cannot confirm that a ¿change in position¿ occurred.I cannot determine the root cause of this event.Causes of early movement of a femoral component are multifactorial.If infection is present this could be a cause.In the absence of infection trauma can cause a change in position due to disruption of the cement mantle or fracture.Surgical technique can also be a factor especially regarding the technique of cementing.The use of a competitor cement brings into question the strength and other characteristics of the product.I have no knowledge of the track record or performance of this particular cement.The performance and clinical outcomes of cemented triathlon total knee arthroplasty is excellent and early femoral loosening is extremely uncommon.This may be due to surgical technique but i cannot be certain of this.I would rule out any causality to the implant itself.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusion: a review of the provided medical records by a clinical consultant indicated: this inquiry concerns a patient who had a primary cemented triathlon ps total knee arthroplasty with a stem extension on the tibia.Less than one year later the patient required revision for a "change in position" of the femoral component.The surgeon elected to use a non-stryker cement.I can only assume that the "change in position" of the femoral component was due to early loosening.There was nothing in the history to suggest trauma.In my experience, early loosening of a cemented femoral component is uncommon.It can occur if infection is present but aseptic loosening is unusual.I cannot confirm that a ¿change in position¿ occurred.I cannot determine the root cause of this event.Causes of early movement of a femoral component are multifactorial.If infection is present this could be a cause.In the absence of infection trauma can cause a change in position due to disruption of the cement mantle or fracture.Surgical technique can also be a factor especially regarding the technique of cementing.The use of a competitor cement brings into question the strength and other characteristics of the product.I have no knowledge of the track record or performance of this particular cement.The performance and clinical outcomes of cemented triathlon total knee arthroplasty is excellent and early femoral loosening is extremely uncommon.This may be due to surgical technique but i cannot be certain of this.I would rule out any causality to the implant itself.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS FEM COMPONENT, CEMENTED
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14540110
MDR Text Key292971622
Report Number0002249697-2022-00782
Device Sequence Number1
Product Code MBH
UDI-Device Identifier07613327050394
UDI-Public07613327050394
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K042993
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/30/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number5515-F-501
Device Catalogue Number5515-F-501
Device Lot NumberEV77XD
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received08/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/14/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age70 YR
Patient SexFemale
Patient Weight67 KG
-
-