• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE MEDIAL CONGRUENT (MC) RIGHT 13 MM HEIGHT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE MEDIAL CONGRUENT (MC) RIGHT 13 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/10/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).42532008302 - tibia cemented 5 degree stemmed right size h - 65132765.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
 
Event Description
It was reported that the poly would not engage and lock into place.A second poly was opened and clicked right in.It was noted that the surgeon tried to insert the first poly without cleaning the tray out first.There was no consequences or impact to the patient.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the provided pictures identified signs of use and showed flared damages.No compression damage was noted from the provided picture.The device history record was reviewed and no discrepancies relevant to the reported event were found.The root cause of the reported issue was attributed to the user not following the instruction in the surgical technique.The surgical technique advises removing the debris in the joint space before implanting.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE MEDIAL CONGRUENT (MC) RIGHT 13 MM HEIGHT
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key14549955
MDR Text Key293373579
Report Number3007963827-2022-00152
Device Sequence Number1
Product Code MBH
UDI-Device Identifier00889024468528
UDI-Public(01)00889024468528(17)250815(10)64767380
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150090
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42522100913
Device Lot Number64767380
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/09/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.
Patient SexMale
-
-