H4: the lot was manufactured between july 07, 2021 to july 08, 2021.H10: the device was received for evaluation containing fluid in the bladder.Visual inspection on the unit via the naked eye was performed and no fluid was observed coming out of the tubing line after the slide clamp was disengaged and the blue winged luer cap was opened.After the tubing line was cut, no evidence of fluid was observed flowing out of the tubing line which indicated blockage at the junction of the tubing and stressmember located at the upper area of the device.The device was disassembled for further examination and the needle located inside the stressmember was found to be misassembled.The needle touched the tubing port which blocked the fluid path and did not allow fluid to flow out of the tubing line.The reported condition was verified.The cause of the condition was due to a manufacturing issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|