It was reported to philips that while the unit was at bench repair, it was found that the paddle cable had a cut on it.The customer requested that the device be returned for bench repair.The device was received by the bench repair service on 16nov2021.Upon evaluation of the device, the reported issue was confirmed and the cause was traced to a faulty paddle cable.Based on the conclusion of the investigation, it was determined that this was a malfunction of the paddle cable.The paddle cable was replaced and the device passed all testing.The device remains at the customer site.There is no indication of a systemic problem.No further investigation or action is warranted.Was the device being used on a patient at the time of the event, including for the purposes of diagnosis? no.Was there any adverse event to the patient or user.If yes, describe? no, the event did not occur during clinical use.If there was an adverse event, did the device cause or contribute to the adverse event, and how? no, there was reportedly no adverse event.
|