Model Number N/A
Device Problems Fluid/Blood Leak (1250); Misassembled (1398); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This complaint has been submitted under (b)(4).Review of the most recent repair record determined the motor speed was unstable and 2 screws were worn.The motor and screws were replaced and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the device closes unevenly, gap after blade application, leaky rails on right side.Upon inspection it was found that the motor speed was unstable and 2 screws were worn.No harm or delay were reported along with the event.No other adverse events were reported along with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key14573433
MDR Text Key293266383
Report Number0001526350-2022-00535
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024375994
UDI-Public(01)00889024375994(11)180108(10)63898207
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/02/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number63898207
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2022
Was the Report Sent to FDA? No
Date Manufacturer Received05/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/12/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-