Model Number IRC10LXO2
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem Partial thickness (Second Degree) Burn (2694)
Event Date 04/30/2022
Event Type  Injury  
Manufacturer Narrative
There was no allegation of malfunction or defect with the concentrator.Based on the information provided, the underlying cause of the event is user error; the patient did not adhere to the device labeling and instructions.The irc10lxo2 user manual states, "do not smoke while using this device.Do not use near open flame or ignition sources.No smoking signs should be prominently displayed.Keep all matches, lighted cigarettes, electronic cigarettes or other sources of ignition out of the room in which this concentrator is located and away from where oxygen is being delivered." in addition, the device itself is prominently labeled regarding the hazard of smoking or exposing the unit to an open flame/ignition source.If additional information becomes available, a supplemental record will be filed.
 
Event Description
A hospice facility stated that a patient was smoking while using an irc10lxo2 concentrator in his home.He sustained second degree burns to the face with swelling and was intubated at the hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVACARE PLATINUM 10 OXYGEN CONCENTRATOR
Type of Device
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
INVACARE FLORIDA
2101 east lake mary blvd
sanford FL 32773
Manufacturer (Section G)
INVACARE FLORIDA
2101 east lake mary blvd
sanford FL 32773
Manufacturer Contact
jason fiest
one invacare way
elyria, OH 44035
8003336900
MDR Report Key14593747
MDR Text Key293276961
Report Number1031452-2022-00030
Device Sequence Number1
Product Code CAW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberIRC10LXO2
Device Catalogue NumberIRC10LXO2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age69 YR
Patient SexMale
Patient Weight117 KG
-
-