• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL 150 - AMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL 150 - AMER Back to Search Results
Model Number 27003
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/10/2022
Event Type  malfunction  
Manufacturer Narrative
The astral device was returned to resmed.Evaluation confirmed the reported complaint.The internal battery was replaced to address the issue.The device was serviced and fully tested before it was returned to the customer.Resmed reference#: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed a battery error message.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
Based on all available evidence and complaint investigations of a similar nature, an investigation determined that the reported complaint was due to an intermittent connection with the battery.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed a battery error message.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - AMER
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
Manufacturer (Section G)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
Manufacturer Contact
michelle hughes
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
MDR Report Key14598757
MDR Text Key293626883
Report Number3004604967-2022-00391
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00619498270033
UDI-Public(01)00619498270033(11)220126(10)1591278
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172875
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27003
Device Catalogue Number27003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2022
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date06/20/2022
Date Manufacturer Received06/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/31/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-