• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK XS PT TEST; PROTHROMBIN TIME TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK XS PT TEST; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 04625315003
Device Problems Inadequate or Insufficient Training (1643); Device Markings/Labelling Problem (2911); Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/10/2022
Event Type  malfunction  
Event Description
The customer state the information on the label of their test strip box does not match the information on the strip vial label.Customer stated the box label states the test strips are lot 56099311 with expiration date 31-mar-2023.The customer stated the test strip vial label states the strips are lot 45392511 with expiration date 30-sep-2020.
 
Manufacturer Narrative
The investigation is ongoing.
 
Manufacturer Narrative
The production process and production dates were investigated.It is not possible for test strip lot 45392511 expiration: 30-sep-2020 to be packed in an outer packaging of test strips that do not expire until 2023.The time difference of 3 years between expiry date of test strip vials and package is impossible.Retention material contains the correct code key.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS PT TEST
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key14602036
MDR Text Key293351431
Report Number1823260-2022-01612
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/31/2023
Device Catalogue Number04625315003
Device Lot Number56099311
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-