Catalog Number 07027796190 |
Device Problem
False Negative Result (1225)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/12/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The qc results from the date of the event were acceptable.A sample from the patient was submitted for further investigation.The investigation is ongoing.
|
|
Event Description
|
There was an allegation of a questionable false-negative rubella igm elecsys result from cobas e 801 module serial number (b)(4).The result was 0.643 (non-reactive).The patient was also tested on a vidas instrument and the result was positive.The same sample was also tested on cobas e601 instrument and the result was non-reactive.The questionable result was reported outside of the laboratory.
|
|
Manufacturer Narrative
|
Clarification was received that the sample was tested on the vidas analyzer on (b)(6) 2022.The same sample was then tested on the cobas e601 instrument on the same day.The next day, (b)(6) 2022, the same sample was tested on the cobas e 801 module.Medwatch field b3 - date of event was updated.A sample from the patient was submitted for investigation.The elecsys rubella igm non-reactive result was confirmed at 0.78 coi.The elecsys rubella igg was non-reactive at 0.210 iu/ml with a data flag.The investigation is ongoing.
|
|
Manufacturer Narrative
|
The provided patient sample was further tested with the platelia rubella igm assay and the mikrogen recomblot rubella igg assay.The platelia rubella igm assay was positive and the recomblot rubella igg was negative with no bands at all.It was determined the rubella igg negativity and the positive result in the platelia rubella igm do not exclude a true positive rubella igm for this sample.The elecsys rubella igm was enhanced and found close to the grey zone.The elecsys rubella igm assay performs within specification.
|
|
Search Alerts/Recalls
|