There was an allegation of questionable high ise indirect gen results from the cobas pro ise analytical unit that were reported outside of the laboratory and were later corrected.Only data for one sample that was not reported outside of the laboratory was provided.The initial sodium result was 146 mmol/l.As the qc results were high, the sample was repeated on another analyzer and the result was 138 mmol/l, which was believed to be correct.The electrode lot number and expiration date were requested but were not provided.
|
The field service representative found there was an issue with the chloride electrode.He checked the ise flow path, calibration, and qc.He replaced the chloride electrode.Repeated ise calibration and qc with results within range.Sodium calibration was ok, with no alarms.Upon review of control data, sodium controls were outside of range on (b)(6) 2022 and (b)(6) 2022.Multiple sample probe aspiration alarms occurred on the day of the event.The investigation determined the service actions resolved the issue.Medwatch fields d1-d4, g1, and g4 were updated.
|