Model Number N/A |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Bone Fracture(s) (1870); Osteopenia/ Osteoporosis (2651); Muscle/Tendon Damage (4532)
|
Event Date 02/27/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Report source: hartford j.M., graw b.P, frosch d.L.(2022) reduced incidence of perioperative periprosthetic fractures using hybrid rasp-impaction broaching over impaction broaching when using the direct anterior approach for total hip arthroplasty.Arthroplasty today vol.15 (2022) 75-80 https://www.Arthroplastytoday.Org/article/s2352-3441(22)00069-3/fulltext.Customer has indicated that the product will not be returned to zimmer biomet for investigation, product location unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported patient experienced avulsion fractures of the greater trochanter.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: g3; h2; h3; h6.H6 component code: mechanical - (g04) stem.Reported event was unable to be confirmed.Device history record (dhr) review was unable to be performed as the item and lot number of the device involved in the event is unknown.Root cause could not be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional information on the reported event.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Additional information does not change the root cause of the previous investigation.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported patient experienced avulsion fractures of the greater trochanter.Surgeon believed osteoporosis, micro fracture, excessive forces from the edges of the shoulder of the implant and broach caused the avulsion fracture.Patient used protected weight bearing to address the fracture.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Search Alerts/Recalls
|