Model Number 26280
Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/18/2022
Event Type  Injury  
Manufacturer Narrative
Age at time of event - 18 years or older.
 
Event Description
It was reported that stent profile problem occurred.The 70% target lesion was located in the non-tortuous and non-calcified left innominate vein.A 14x60/9fr uni plus 75cm wallstent endoprosthesis was advanced for treatment.However, it was found out that there might be a flaw on the tip part of the stent while it was released.The physician used a balloon to solve the issue and the stent was successfully implanted.No patient complications were reported and the patient status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLSTENT ENDOPROSTHESIS
Type of Device
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14633689
MDR Text Key293561815
Report Number2134265-2022-05961
Device Sequence Number1
Product Code MAF
UDI-Device Identifier08714729204008
UDI-Public08714729204008
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K152842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/17/2023
Device Model Number26280
Device Catalogue Number26280
Device Lot Number0026826870
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/17/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-