WRIGHT MEDICAL TECHNOLOGY INC INFINITY TIBIAL TRAY SIZE 3 TOTAL ANKLE SYSTEM; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Catalog Number 33650003 |
Device Problem
Loss of Osseointegration (2408)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 05/13/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed. should additional information become available, it will be provided on a supplemental report.Device remains implanted in patient.
|
|
Event Description
|
It was reported that the patient underwent a total ankle replacement.Allegedly, the patient may need to undergo a revision surgery for aseptic tibial tray loosening.
|
|
Manufacturer Narrative
|
The reported event could be confirmed since images of ct scans matches the alleged failure.There are many clinical factors that can affect the results of any surgery, such as surgical technique, pre-operative and post-operative care, the implant, patient pathology and daily activity.A medical professional reviewed the received information and noted the following: "i agree upon your conclusion of loosening (and tilting) of the tibial tray of this tar.In the information provided, there are no other obvious signs of patient problems that can be discerned." more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
|
Event Description
|
It was reported that the patient underwent a total ankle replacement.Allegedly, the patient may need to undergo a revision surgery for aseptic tibial tray loosening.
|
|