Catalog Number 2C2111K |
Device Problem
Material Rupture (1546)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/11/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Initial reporter phone no.(b)(6).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the balloon of a small volume intermate burst during filling.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.A visual inspection performed using the naked eye found the bladder had been ruptured.The ruptured bladder was examined for signs of abnormality that may have potentially caused the rupture problem.No signs of abnormality were found.The reported condition was verified.The cause of the condition could not be determined; however the most probable cause is manufacturing related.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|