• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON GRIESHABER AG BACKFLUSH HANDLE DSP; INSTRUMENT, VITREOUS ASPIRATION AND CUTTING, AC-POWERED, ACCESSORY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON GRIESHABER AG BACKFLUSH HANDLE DSP; INSTRUMENT, VITREOUS ASPIRATION AND CUTTING, AC-POWERED, ACCESSORY Back to Search Results
Catalog Number 337.88
Device Problem Disconnection (1171)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 05/16/2022
Event Type  Injury  
Event Description
A physician reported that the tip of an ophthalmic backflush device fell off in the patient's eye during a vitrectomy procedure.The eye is filled with silicone oil.The physician indicated that the tip might come out when the silicone oil would be removed from the eye.The current patient prognosis was good.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
No sample has been received at manufacturing for evaluation.A review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The root cause cannot be identified conclusively because no picture or sample has been received.This complaint has been reviewed and future data will be monitored for evidence of adverse trending and further action will be taken, as appropriate.At a minimum, this will include completing reviews of complaint class report levels on a monthly basis.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BACKFLUSH HANDLE DSP
Type of Device
INSTRUMENT, VITREOUS ASPIRATION AND CUTTING, AC-POWERED, ACCESSORY
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14654915
MDR Text Key293724288
Report Number3003398873-2022-00038
Device Sequence Number1
Product Code HQE
UDI-Device Identifier07612717071414
UDI-Public07612717071414
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
K884043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2023
Device Catalogue Number337.88
Device Lot NumberF195889
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-