Explant due to regurgitation and shortness of breath was reported.The investigation found that all three leaflets contained tears.There was also focal fibrous pannus on the outflow surface of leaflet 3.No acute inflammation, significant calcification, or stent deformation was found.Information from the field indicated that it was noted at explant that the valve had fused with the aorta the device history record was reviewed to ensure each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.This was inclusive of a review of the manufacturing videos, which contained no evidence of anomalies during functional inspection.In the absence of any calcification or evidence for infection, the reported event is consistent with a non-calcific leaflet tear.A non-calcific leaflet tear is a form of structural valve deterioration (svd), which is a well-known complication from valve replacement surgery.A non-calcific leaflet tear is commonly attributed to increased operational leaflet stress but may also be related to biological factors which result in tissue degeneration characterized by loss of collagen.In this case, histological evaluation did not demonstrate loss of collagen at the tear site, and the cause of the leaflet tears could not be conclusively determined.However, at the time of valve explant there was evidence of fusion of a stent post to the aortic wall, which may be indicative of a narrow aortic sinus relative to the implanted valve size.A narrow aortic sinus may increase the interaction between the aortic wall and stent posts and has the potential to result in abrasion of the leaflet tissue along the stent post including the formation of pannus, which was noted on the explanted valve.The interaction between the stent posts and the aortic wall has the potential to induce increased stress on adjacent leaflets and create an unbalanced stress relief distribution between all leaflets during coaptation, which in this case may have led to the observed leaflet tears and reduced durability.There is no indication of a product quality issue with regards to manufacture, design, or labeling.Na.
|