Model Number RONYX35015UX |
Device Problem
Component Missing (2306)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/06/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
During a procedure one resolute onyx rx coronary drug eluting stent was attempted to be used.Damage was not noted to the packaging.There were issues noted when removing the device from the hoop.The device was inspected with issues noted.Negative prep was not performed.It was reported that there was no stent on the delivery balloon.The stent was noted to be missing from the balloon catheter upon removal of the device from the hoop.The physician stated that no stent was on the balloon prior to use but the balloon was used on the patient.No patient injury reported.
|
|
Manufacturer Narrative
|
Product analysis summary: kinks evident on the hypotube.A kink was evident on the distal shaft and core wire.The stent was not present on the balloon and did not return for analysis.The stent was not present on the balloon and did not return for analysis.Crimp impressions were not visable on the balloon.The balloon returned deflated.Deformation evident to the distal tip.The inner lumen could not be verified with a 0.015 inch mandrel most likely due to hardened blood in the guidewire lumen.No other damage evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|