• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIRGAS THERAPEUTICS LLC ISPAN GAS TANK REGULATOR; INTRAOCULAR GAS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIRGAS THERAPEUTICS LLC ISPAN GAS TANK REGULATOR; INTRAOCULAR GAS Back to Search Results
Catalog Number 8065797303
Device Problems Gas Output Problem (1266); Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/19/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported that during a vitrectomy procedure an ophthalmic regulator knob was fallen off and it could not dispense the perfluropropane (c3f8).The procedure was completed using silicone oil for the patient.There was no harm to the patient.
 
Manufacturer Narrative
Additional information was provided in sections h.6 and h.10.With the information provided, the customers¿ reported event was not able to be confirmed.A check of the batch production record for this lot, showed no unusual manufacturing issues.A check of the complaint records showed (b)(4) complaints against the same lot number.A check of confirmed complaints showed (b)(4) similar complaints.There was no sample returned for this investigation.With no additional, related information provided, the customer reported event was not confirmed.Based upon the information obtained, the root cause cannot be determined conclusively.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISPAN GAS TANK REGULATOR
Type of Device
INTRAOCULAR GAS
Manufacturer (Section D)
AIRGAS THERAPEUTICS LLC
6141 easton road
p.o. box 310
plumsteadville PA 18949
Manufacturer (Section G)
AIRGAS THERAPEUTICS LLC
6141 easton road
p.o. box 310
plumsteadville PA 18949
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14679394
MDR Text Key294410122
Report Number1610287-2022-00043
Device Sequence Number1
Product Code LPO
UDI-Device Identifier00380657973033
UDI-Public00380657973033
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P900066
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 10/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065797303
Device Lot Number104323
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-