It was reported that this patient went through a revision procedure due to oversensing without pacing inhibition, loss of capture, and high out-of-range pacing impedance for both unipolar and bipolar configurations, in both the right atrial (ra) and right ventricular leads.Reportedly, impedance was jumpy/irregular on the atrial channel, while consistently high on the ventricular channel.The patient suffered sinus bradycardia as a result of these issues.Upon the revision procedure, both leads were tested in both configurations in the pacing system analyzer, and no abnormalities were found.Subsequently, the physician elected to only replace the pacemaker, because when the leads were connected to the new pacemaker, no issues were observed again.However, it was indicated that these issues were related to poor terminal pin to header connection.This ra lead remains in service and no additional adverse patient effects were reported.
|