Model Number OER-PRO
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device has not been returned to olympus for evaluation.It was noted, the customer reached out to the field service engineer (fse) to resolve the issue.During troubleshooting, the central processing unit (cpu) was reset, the lcg was replaced but the light stayed on.Customer was advised to contact the fse once the aldohol was replaced again, to see if the light goes out.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported to olympus, the liquid chemical germicide (lcg) light of the endoscope reprocessor stayed on even after it has been reloaded.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Sections e2 and e3 were inadvertently selected in error.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and since the device was not returned for evaluation, the definitive root cause of the disinfectant replacement indicator not disappearing after replacing the disinfectant could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14691020
MDR Text Key302360067
Report Number8010047-2022-10031
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 10/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/07/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-