The customer reported to olympus during preparation for a therapeutic steroid injection for subglottic stenosis procedure, multiple devices failed and the fluid did not flow out the needle when deployed outside of the patient with the injector in it's resting coiled position.These events will be submitted separately.The customer also reported, the fourth subject device failed during an unspecified procedure while the patient was awake with the endoscope in their nasopharynx.The subject device was tested prior to use and there were no issues.The customer also reported visibility in the patient was diminished due to secretions.The patient's tolerance was decreasing rapidly with gagging and coughing and the procedure was not able to be completed.This event includes 4 reports: (b)(6): for device failure inside patient.(b)(6): for device failure before use with patient.(b)(6): for device failure before use with patient.(b)(6): for device failure before use with patient.This report is 1 of 4 for (b)(6):the device that failed inside the patient.
|
This supplemental report is being submitted to provide the results of the legal manufacturer¿s investigation and the results of the device history records (dhr) review.New information added to the following fields: h6, h10.H4: february 2020.The dhr for this device were reviewed and all records indicated the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.A definitive root cause was not identified.Based on the available information, the legal manufacturer determined the probable cause of the failure is likely due to: the needle was extended because of the great friction between the outer tube and the needle.The instruction manual contains the following descriptions, and it warns against this event.(rk1556 rev.01) straighten out the instrument before inspecting it.The instrument can be damaged if it is coiled while the handle is operated.Operate the slider slowly, otherwise the tube could buckle.When inserting the instrument into the endoscope, retract the needle into the sheath, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.Insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.Stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.Olympus will continue to monitor field performance for this device.
|