Catalog Number 8416400
Device Problem Fire (1245)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/18/2022
Event Type  malfunction  
Event Description
It was reported that: "drager infinity v500 ventilator using a flex care heater with disposable, single use temperature probe was in use.The respiratory therapist noted an orange glow and sizzle sound followed by a one single "chirp" sound made by the ventilator equipment.Instantly, flames melted the plastic hoses which connected the patient to the ventilator machinery.The patient was immediately moved to safety, unharmed, and ventilation was provided by oxygen tank and ambu-bagging.The ventilator equipment fire was extinguished quickly before it could spread beyond the single room.Several healthcare workers were evaluated in the emergency department for potential smoke inhalation and dust from use of fire extinguisher equipment.All involved employees were discharged home from the emergency department." there was no injuy reported.The device is presently sequestered and not available to draeger for investigation.
 
Manufacturer Narrative
The investigation has just started; results will be provided in a follow-up report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY ACS WORKSTATION CC
Type of Device
VENTILATORS, INTENSIVE CARE
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
MDR Report Key14731226
MDR Text Key294942491
Report Number9611500-2022-00148
Device Sequence Number1
Product Code CBK
UDI-Device Identifier04048675042266
UDI-Public(01)04048675042266(11)201208(17)210930(93)8416400-21
Combination Product (y/n)N
PMA/PMN Number
K093633
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8416400
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/18/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/30/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-