Model Number 37800 |
Device Problems
Impedance Problem (2950); Material Deformation (2976)
|
Patient Problems
Nausea (1970); Pain (1994)
|
Event Date 05/23/2022 |
Event Type
Injury
|
Event Description
|
Information was received from a healthcare provider regarding a patient who was implanted with an implantable neurostimulator (ins) for gastrointestinal/pelvic floor.It was reported that patient reported to the hospital with increased nausea and pain.Impedance was out of range 20,000.A x-ray showed the leads were in knots within the abdomen.The hcp plans on removing both the leads and battery and replacing them.
|
|
Manufacturer Narrative
|
Concomitant medical products: product id: 4351-35, serial# (b)(4), implanted: (b)(6) 2019, explanted: product type: lead.Product id: 4351-35, serial# (b)(4), implanted: (b)(6) 2019, explanted: product type: lead.Other relevant device(s) are: product id: 4351-35, serial/lot #: (b)(4), ubd: 29-may-2021, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from a manufacturer representative (rep).The rep reported that patient had a new battery on july 8th.
|
|
Manufacturer Narrative
|
Continuation of d10: product id neu_unknown lot# serial# unknown implanted: explanted: product type unknown product id 4351-35 lot# serial# (b)(6), implanted: (b)(6)2019, explanted: product type lead product id 4351-35 lot# serial# (b)(6), implanted: (b)(6)2019, explanted: product type lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|