Model Number V60
Device Problem Volume Accuracy Problem (1675)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/08/2022
Event Type  malfunction  
Manufacturer Narrative
Reporting address line 1: (b)(6).Reporting address state: (b)(6).Reporting address postal: (b)(6).Reporting institution phone #: (b)(6).Reporter phone #: (b)(6).
 
Event Description
It was reported that the ventilators tidal volume was not working.Reportedly, with a 21% fio2 setting, the unit was showing a very low delivered tidal volume (>20ml).However, with a 100% fio2 setting the unit was showing normal delivered tidal volume and intermittent patient disconnect alarms.There was no patient involvement.Further information has been requested.If additional information becomes available at a later date, a supplemental report will be submitted.
 
Manufacturer Narrative
The repair for this unit cannot be conducted at this time due to the part(s) being on back order.The material has already been requested and an order for the part(s) was placed to conduct the repair of this unit.The material ordered flow sensor assembly aligns with the recommended repair of the reported malfunction per the service manual.When the parts become available the repairs will be conducted.If new information is received and suggests that the device has additional malfunctions, the record will be reopened, and an investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
kimberly shelly
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key14733957
MDR Text Key301917464
Report Number2031642-2022-01631
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053614
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received09/27/2022
Date Device Manufactured10/23/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-