RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Volume Accuracy Problem (1675)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/08/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Reporting address line 1: (b)(6).Reporting address state: (b)(6).Reporting address postal: (b)(6).Reporting institution phone #: (b)(6).Reporter phone #: (b)(6).
|
|
Event Description
|
It was reported that the ventilators tidal volume was not working.Reportedly, with a 21% fio2 setting, the unit was showing a very low delivered tidal volume (>20ml).However, with a 100% fio2 setting the unit was showing normal delivered tidal volume and intermittent patient disconnect alarms.There was no patient involvement.Further information has been requested.If additional information becomes available at a later date, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
The repair for this unit cannot be conducted at this time due to the part(s) being on back order.The material has already been requested and an order for the part(s) was placed to conduct the repair of this unit.The material ordered flow sensor assembly aligns with the recommended repair of the reported malfunction per the service manual.When the parts become available the repairs will be conducted.If new information is received and suggests that the device has additional malfunctions, the record will be reopened, and an investigation will be performed.
|
|