Catalog Number UNKNOWN
Device Problem Biocompatibility (2886)
Patient Problems Ambulation Difficulties (2544); Metal Related Pathology (4530); Swelling/ Edema (4577)
Event Date 10/28/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that, after a bhr resurfacing construct had been implanted on the patient¿s left hip on (b)(6) 2009, the patient experienced hip pain, swelling, metallosis, difficulty walking, soft tissue damage and metal-on-metal reaction within the left hip joint.A revision surgery was conducted on (b)(6) 2021 to address this adverse event.The identity of the explanted devices and the ones implanted in exchange are not known.Patient¿s current health status is not known.
 
Manufacturer Narrative
H3, h6.It was reported that, after a bhr resurfacing construct had been implanted on the patient¿s left hip on (b)(6) 2009, the patient experienced hip pain, swelling, metallosis, difficulty walking, soft tissue damage and metal-on-metal reaction within the left hip joint.A revision surgery was conducted on 28-oct-2021 to address this adverse event.The identity of the explanted devices and the ones implanted in exchange are not known, and the parts, used in treatment, were not available for analysis.As no device part and batch numbers were provided for investigation, a complaint history review, manufacturing record review, device labelling / ifu review could not be performed.If more information is received, this investigation will be reopened.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions are required.A risk management review was performed for the acetabular cup and femoral head.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.Based on the reported symptoms it cannot be concluded that the events/clinical reactions (soft tissue damage and metal-on-metal reaction) were associated with a mal-performance of the implant.The patient impact cannot be determined at this time.No further clinical assessment is warranted at this time.Without return of the actual devices or further information we cannot further investigate the details supplied in this complaint, our investigation remains inconclusive, and a definitive root cause cannot be determined.Based on the limited information provided we are unable to speculate on specific factors known to contribute to the alleged fault.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.H11: corrected information in h6 (health effect - clinical code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN BIRMINGHAM HIP RESURFACING (BHR) HEAD
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14741430
MDR Text Key294479274
Report Number3005975929-2022-00401
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention; Other;
Patient SexMale
-
-