Model Number 11823
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/23/2022
Event Type  malfunction  
Manufacturer Narrative
Pending follow up information.A review of the device history record, which includes verification of all steps in the manufacturing of the us catheter kit, verification of all final testing performed by/on the us catheter kit, and packaging for subject us catheter kit was performed.The review did not identify any non-conformance, issues or capas associated with us catheter kit function.Device was not returned for additional evaluation and investigation.Per the instructions for use of the device, catheter kinking is a known possible risk of use of the device.If additional information becomes available, a supplemental mdr will be sent.Internal complaint number: (b)(4).
 
Event Description
Information was received reporting that a dye study was performed confirming that the patient's catheter was kinked.Reporter confirmed that no catheter revision surgery was performed because of the patient's health.
 
Manufacturer Narrative
Additional communication with representative confirmed that the cause of the catheter kink was unknown.Internal complaint number: complaint-(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRATHECAL CATHETER
Type of Device
INTRATHECAL CATHETER
Manufacturer (Section D)
FLOWONIX MEDICAL, INC.
500 international drive
suite 200
mount olive NJ 07828
Manufacturer (Section G)
FLOWONIX MEDICAL, INC.
120 forbes blvd
suite 170
mount olive MA 07828
Manufacturer Contact
james bennett
500 international drive
suite 200
mount olive, NJ 07828
9734269229
MDR Report Key14775529
MDR Text Key294540340
Report Number3010079947-2022-00119
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00810335020099
UDI-Public(01)00810335020099(17)220909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P080012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/09/2022
Device Model Number11823
Device Catalogue Number11823
Device Lot Number26908
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/09/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-