• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number RA002-5555SL
Device Problems Unstable (1667); Naturally Worn (2988); Scratched Material (3020)
Patient Problem Failure of Implant (1924)
Event Type  malfunction  
Event Description
An investigation report from an implant retrieval center was received and reported that a revision procedure was performed due to pullout.During examination of the rod, the retrieval center found discoloration, wear, galling, scratches, and other surface damage.There was no further reported patient impact and no patient or user injury was reported.No additional information is available.Report 1 of 2.
 
Manufacturer Narrative
No product has been returned for evaluation.Evaluation was performed by (b)(4).The root cause is unable to be determined at this time.A review of the device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.If any additional information is provided, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key14793944
MDR Text Key302256404
Report Number3006179046-2022-00222
Device Sequence Number1
Product Code PGN
UDI-Device Identifier00856719002190
UDI-Public856719002190
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K160352
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRA002-5555SL
Device Lot NumberA170207-05
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/07/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age7 YR
Patient SexMale
Patient Weight20 KG
-
-