U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 VENT, CHINA OPT: CFLEX, AVAPS
Device Problem Volume Accuracy Problem (1675)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/17/2022
Event Type  malfunction  
Event Description
The customer reported a low tidal volume.It was reported that the device was in use at the time of the event.No patient or user harm reported.The customer observes the error message displayed on the display.It was recommended to have onsite service performed.Further information has been requested.If additional information becomes available at a later date, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(6).
 
Manufacturer Narrative
The key market reported that the issue was caused by water entering the pressure monitoring pipeline.After the service engineer cleared out the water, the device was returned to normal operation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
kimberly shelly
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key14800550
MDR Text Key294847821
Report Number2031642-2022-01685
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838025776
UDI-Public(01)00884838025776
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60 VENT, CHINA OPT: CFLEX, AVAPS
Device Catalogue Number1076716
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received07/24/2022
Date Device Manufactured03/29/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-