• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION JETSTREAM XC ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION JETSTREAM XC ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 45007
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/27/2022
Event Type  malfunction  
Manufacturer Narrative
Device eval by manufacturer: returned product consisted of a jetstream xc-2.4 atherectomy catheter.The device was visually examined for any shaft damage and the functional testing of the device was completed.The device showed damage in the form of buckling/kinks locate 8cm from the tip to proximal 13cm.Functional analysis was performed and revealed that the rotation of the device functioned as designed.Aspiration testing was performed using a 100ml beaker of water.The devices tip is submerged in a beaker of fluid and the device is run for a period of 1 minute.Test results showed that this device did not perform as designed per the test procedure specification sheet withdrawing 0ml of fluid in the 1 minute time frame.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.The complaint was not confirmed for rotational issues; however, evacuation failure was confirmed.
 
Event Description
Reportable based on analysis completed on 02jun2022.It was reported that this device stopped operation during treatment a jetstream xc 2.4mm atherectomy catheter was selected for use.The target lesion was located in the superficial femoral artery (sfa) and the plaque was observed to be very soft and fresh.During treatment of the lesion, the device stopped rotating and working.The procedure was able to be completed successfully using another jetstream without sequela.However, device analysis revealed the device lost aspiration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM XC ATHERECTOMY CATHETER
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14802926
MDR Text Key296182964
Report Number2134265-2022-06643
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/05/2024
Device Model Number45007
Device Catalogue Number45007
Device Lot Number0028994161
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/26/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age48 YR
-
-