The consumer reported a (b)(6) result with the alere determine hiv 1/2 ag/ab combo 25t test performed on 14oct2021.Confirmation testing was performed using biorad hiv combo ag/ab eia, generated a (b)(6) result.The customer stated there was a possible c-section due to the test results.Additionally, the customer confirmed there was a live birth.
|
Testing was performed at abbott diagnostics scarborough, inc.On retained kit lot 161479 with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for test kit part number 7d2648 / lot 161479, test base part number 10732998 / lot 161479.The lot met the required release specifications.(b)(4).Abbott diagnostics scarborough was unable to determine the exact root cause of the reported issue.However a possible assignable root cause is sample interference or cross contamination.
|