• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ALERE DETERMINE HIV 1/2 AG/AB COMBO 25T; IVD FOR HIV 1/2 AG/AB

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ALERE DETERMINE HIV 1/2 AG/AB COMBO 25T; IVD FOR HIV 1/2 AG/AB Back to Search Results
Catalog Number 7D2648
Device Problem False Positive Result (1227)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/14/2021
Event Type  Injury  
Event Description
The consumer reported a (b)(6) result with the alere determine hiv 1/2 ag/ab combo 25t test performed on 14oct2021.Confirmation testing was performed using biorad hiv combo ag/ab eia, generated a (b)(6) result.The customer stated there was a possible c-section due to the test results.Additionally, the customer confirmed there was a live birth.
 
Manufacturer Narrative
Testing was performed at abbott diagnostics scarborough, inc.On retained kit lot 161479 with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for test kit part number 7d2648 / lot 161479, test base part number 10732998 / lot 161479.The lot met the required release specifications.(b)(4).Abbott diagnostics scarborough was unable to determine the exact root cause of the reported issue.However a possible assignable root cause is sample interference or cross contamination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALERE DETERMINE HIV 1/2 AG/AB COMBO 25T
Type of Device
IVD FOR HIV 1/2 AG/AB
Manufacturer (Section D)
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
10 southgate road
scarborough ME 04074
Manufacturer Contact
rachel blackwell
10 southgate road
scarborough, ME 04074
6613888803
MDR Report Key14804578
MDR Text Key295040792
Report Number1221359-2022-03300
Device Sequence Number1
Product Code MZF
UDI-Device Identifier10811877010293
UDI-Public011081187701029310161479
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BP120037-0
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/24/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number7D2648
Device Lot Number161479
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/15/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age30 YR
Patient SexFemale
Patient EthnicityHispanic
-
-