Device analysis: one xen 45 gts injector was received.The needle cover covered the needle, the needle plug and gel stent were not returned.The cam lock was received detached from the injector.The slider of the injector was not in the start position, but at the base of the injector and the bevel selector in the middle position with the needle retracted.Multiple points of damage were observed along both slider rails on the injector body.A functional test was performed.The slider knob was able to slide the full distance of the travel length in each of the three bevel selector positions with some resistance/interruption at the portion where the rails are damaged.The functional testing showed that the pusher rod which pushes the gel stent out of the needle performed correctly and the needle retracted properly as the slider knob was advanced to the end of the travel distance.The reported complaint of broken/damaged component and blocked slider is confirmed since damage was observed to the slider path on the injector body and slider resistance was observed during functional testing.The resistance observed during functional testing is likely due to the injector damage observed.It was found damage along the pathway of the slider where the cam lock was removed.The complaint issue was able to be recreated (i.E., obstructing the pathway) when twisting the cam lock to remove it rather than "grasping and pulling up." it¿s possible that the user technique in removing the cam lock was a contributing factor, however, there is insufficient information to make a definitive conclusion.Since there is no significant trend in this issue, no further investigation is required.However, it will continue to be tracked and trended.
|