Catalog Number 08D06-77
Device Problem False Negative Result (1225)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/18/2022
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.This report is being filed on an international product, list number 8d06-39 has a similar product distributed in the us, list number 8d06-31/-41.
 
Event Description
The customer observed false nonreactive architect syphilis results generated for a 70 year old female patient.The following data was provided: (b)(6) 2022 initial result = 0.89 s/co, repeat = 0.95 s/co, repeat with another method (fluorescence) = 19.21 (positive), tppa = positive.No impact to patient management was reported.
 
Manufacturer Narrative
The complaint investigation for falsely reactive architect syphilis results included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, device history review, and in house testing was completed.Return testing was not completed as returns were not available.Trending review did not identify any trends for the issue for the product.Device history record review did not identify any non-conformances or deviations with lot 32478be01 and the complaint issue.In-house specificity testing was completed with complaint lot number 32478be01.All controls met specifications and no false reactive results were obtained, indicating that the specificity performance is not negatively impacted.Labeling was reviewed and sufficiently addresses the customer's issue.Based on our investigation, no systemic issue or deficiency with the architect syphilis reagent for lot 32478be01 was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT SYPHILIS TP REAGENT KIT
Type of Device
ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key14850498
MDR Text Key299093891
Report Number3002809144-2022-00224
Device Sequence Number1
Product Code LIP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K153730
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/06/2022
Device Catalogue Number08D06-77
Device Lot Number32478BE01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/22/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ARC I2000SR INST, 03M74-02, (B)(6)
Patient Age70 YR
Patient SexFemale
-
-