• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2022
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in miami, florida.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The failure was traced back to a defective temperature probe.The part was replaced and the problem could be solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova received report that a heater cooler 3t gave an error code associated to discrepancy between measured temperatures between safety and control temperature sensors on the patient side during service.There was no patient injury.
 
Manufacturer Narrative
H.10: based on device history review, it can be highlighted that it was installed since 2011 and no other similar issue occurred.Based on available information and service activity results, the root cause of the reported issue can be traced back to an inaccurate measurement between the control and safety system due to a drifted sensor.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key14874218
MDR Text Key295340266
Report Number9611109-2022-00320
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/05/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-