MAUDE Adverse Event Report: SYNTHES GMBH 2.0MM DRILL BIT W/DEPTH MARK QC/110MM; BIT, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 2.0MM DRILL BIT W/DEPTH MARK QC/110MM; BIT, SURGICAL Back to Search Results
Model Number 310.534
Device Problem Failure to Cut (2587)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/05/2022
Event Type  malfunction  
Event Description
Device report from colombia reports an event as follows: it was reported that during an unknown procedure on (b)(6) 2022, two drill bits were found to be unable to cut.The procedure was completed successfully with no surgical delay.This report is for a 2.0mm drill bit w/depth mark qc/110mm.This is report 2 of 2 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: investigation summary
=
> the complaint device was not received for investigation.A photo investigation was performed based on the provided product photographs.Visual analysis of the photo revealed that there was no damage or defects with 310.534, drill bit ø2 w/marking l110/85 2flute, dullness cannot be confirmed with the evidence provided.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected and released to approved specifications.The overall complaint was not confirmed for 310.534, drill bit ø2 w/marking l110/85 2flute.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventive action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review
=
> a manufacturing record evaluation was performed for the part # 310.534, lot # l281344; and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0MM DRILL BIT W/DEPTH MARK QC/110MM
Type of Device
BIT, SURGICAL
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BETTLACH
muracherstrasse 3
bettlach 2544
SZ   2544
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 
3035526892
MDR Report Key14878804
MDR Text Key302602087
Report Number8030965-2022-04513
Device Sequence Number1
Product Code GFG
UDI-Device Identifier10886982186751
UDI-Public(01)10886982186751
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number310.534
Device Catalogue Number310.534
Device Lot NumberL281344
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-