It was reported the procedure was performed in the distal right coronary artery.The dragonfly optis imaging catheter was inserted, but it was observed that air was in the catheter.When attempting to remove the device, it became entangled with the non-abbott guidewire.Both the guidewire and catheter were able to be removed as a single unit.Another device was used to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no indication of a lot specific product quality issue.Based on the information provided, the reported difficulties appear to be due to operational context.It is likely that the reported interaction with the guidewire, as well as the condition of the guidewire (kinks/bends), caused the difficulty while attempting to remove the catheter; however, this could not be confirmed.The reported aspiration issue could be due to the catheter being damaged, or loose/inadequate fitment of the syringe onto the luer fitting; however, this could also not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|