Model Number A710962
Device Problem Difficult to Flush (1251)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/07/2022
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 04/2021).
 
Event Description
It was reported that approximately one year and five months post port placement, the septum was allegedly could not be located due to blood clot and had difficulty in flushing.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one powerflow implantable port with an attached catheter was returned for evaluation.Functional, gross visual, tactile and microscopic visual evaluations were performed.A complete circumferential break was noted on the distal end of the catheter attached.However, striations were noted throughout the surface.The investigation is inconclusive for the reported difficult to flush, as the exact circumstances at the time of the reported event cannot be verified, and the reported event could not be reproduced in the lab.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiry date: 04/2021), h11: h6 (method, result, conclusion) h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event Description
It was reported that approximately one year and five months post port placement, the septum was allegedly could not be located due to blood clot and had difficulty in flushing.It was further reported that the device was removed.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERFLOW APHERESIS IV IMPLANTABLE PORT, CHRONOFLEX, 9.6F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14883402
MDR Text Key298027608
Report Number3006260740-2022-02492
Device Sequence Number1
Product Code PTD
UDI-Device Identifier00801741129438
UDI-Public(01)00801741129438
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA710962
Device Catalogue NumberA710962
Device Lot NumberREDX0823
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-