Model Number CI-1601-05 |
Device Problems
Adverse Event Without Identified Device or Use Problem (2993); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
Patient Problems
Wound Dehiscence (1154); Erosion (1750); Failure of Implant (1924)
|
Event Date 07/01/2022 |
Event Type
Injury
|
Event Description
|
The recipient reportedly experienced a skin dehiscence at the implant site.On (b)(6) 2022, the recipient underwent skin and muscle flap surgery.The recipient was given stitches and medication (type unknown) but the issue did not resolve.Revision surgery is scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient has reportedly healed.The recipient will not be reimplanted at this time.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed sliced silicone overmold on the top cover, as well as a severed electrode.These anomalies are believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock was verified.The condition of the electrode prevented and electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|