The subject device was returned to an olympus service center for evaluation.During inspection and testing, service observed the flow display was abnormal due to a defective manifold unit.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final.It has been over 8 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, the cause of the phenomenon was determined to be a faulty manifold unit causing the flow to not display normally.No further information could be obtained, and the cause of faulty manifold unit could not be identified.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Olympus will continue to monitor field performance for this device.
|