Model Number CI-1500-01 |
Device Problems
Mechanical Problem (1384); Misconnection (1399); Therapeutic or Diagnostic Output Failure (3023)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 09/29/2022 |
Event Type
malfunction
|
Event Description
|
The patient reportedly experienced loss of lock.Programming adjustments were made, and external equipment was exchanged, however, the issue did not resolve.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Revision surgery will be scheduled.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
The recipient's device was reportedly explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode array was severed.This is believed to have occurred during revision surgery.In addition, this inspection revealed a dent on the bottom cover, as well as a crack on the seam weld.The device passed the photographic imaging inspection.System lock could not be obtained at any spacing.The no lock and electrode condition prevented an electrical test from being performed.The device passed the electrical tests performed.The residual gas analysis revealed loss of hermeticity caused by the damage to the seam weld.The failure of this device is attributed to excessive moisture through a gross leak at the seam weld.This ultimately caused the device to cease functioning.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|