OLYMPUS MEDICAL SYSTEMS CORP. THUNDERBEAT 5 MM, 20 CM, FRONT-ACTUATED GRIP TYPE S; ULTRASONIC SURGICAL DEVICE Back to Search Results
Model Number TB-0520FCS
Device Problems Break (1069); Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/06/2022
Event Type  malfunction  
Manufacturer Narrative
The device has been returned but the device evaluation is not yet completed.As such, a definitive root cause of the reported complaint cannot be determined at this time.Supplemental report(s) will be filed as any relevant new information becomes available.
 
Event Description
As reported for this event by the customer, during a whipple procedure when the device probe was being cleaned, the device probe was found to have a crack and be broken.The probe tip was still attached to the device.The procedure was completed with a new similar device without any delay.There is no reported harm or adverse impact to the patient.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.The device was returned and an evaluation was completed for it.During inspection, olympus confirmed the reported event, the probe was found to be broken.Additionally, scratches on the broken surface of the probe were observed.It was also discovered that the coating of the scratched area was peeled off.It was discovered that the scratches were progressing from the cracks, and the coating of the grasping section was partially peeled off; however, these defects are not considered severe enough to cause a potential adverse event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Although a definitive root cause cannot be identified, the following step-by-step scenario likely caused the event: 1) during the output activation in seal & cut mode, the probe was contacting hard tissue, metal objects or surgical instruments.2) due to ultrasonic vibration, the coating of the probe peeled off.Also, scratches were generated.3) a force to activate the output in seal & cut mode, or a force to grasp the tissue was applied to the probe.Therefore, cracks were generated at the scratched area.4) a force was applied to the probe causing it to break.This following information is included in the instructions for use (ifu): ¿the thunderbeat instrument should be used for soft tissue.Do not activate output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator, forceps, and others).Otherwise, it may cause the probe tip to be scratched or come into direct contact with the metal area of the grasping section as the heat generated by the friction between the hard object and the probe tip could cause wear/deforming/splitting/protruding/partial separating of the tissue pad.In turn, the probe may break before displaying an error window or generating an alarm tone.¿ ¿do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities.¿ ¿if the grasping section, metal-exposed area around it or the probe tip gets sticked tissue during treatment, wipe it with a soft object such as a piece of gauze or a brush.Do not attempt to scrape it with a sharp object such as a scalpel or the tip of tweezers.Otherwise, the grasping section, metal-exposed area around it, the fluorine resin part, a coated surface or the probe tip may be scratched and damaged, which may lead to fall-off of the damaged part into the body cavity or burns of the tissue by a high-frequency leak current output due to destruction of the insulation structure.¿ olympus will continue to monitor the performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THUNDERBEAT 5 MM, 20 CM, FRONT-ACTUATED GRIP TYPE S
Type of Device
ULTRASONIC SURGICAL DEVICE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14954834
MDR Text Key295477568
Report Number8010047-2022-11559
Device Sequence Number1
Product Code GEI
UDI-Device Identifier04953170383588
UDI-Public04953170383588
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K211838
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/08/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTB-0520FCS
Device Lot Number16K28
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/27/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-