BD (SUZHOU) BD INTIMA II¿ IV CATHETER PRN ADAPTER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383019
Device Problems Burst Container or Vessel (1074); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: a device history review was conducted for lot number: 1230634.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.Function test (45psi leakage test) is performed on retained sample of the complaint lot, the result is qualified.Based on the description of the event, our engineers were able to determine that the most likely root cause for this event is the forcing of fluid through the device during injection.The bd pegasus is an infusion only device and is not rated for high pressure injections.
 
Event Description
It was reported that the bd intima ii¿ iv catheter prn adapter the needle burst tube and leakage connection between the one-time connecting pipe and the indwelling needle.The following information was provided by the initial reporter, translated from chinese: ¿the patient went to our hospital for enhanced ct examination on (b)(6) 2022.Contrast agent was injected and the needle burst tube was indwelled during the injection.Leakage occurred at the connection between the one-time connecting pipe and the indwelling needle".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INTIMA II¿ IV CATHETER PRN ADAPTER
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
Manufacturer (Section G)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14979800
MDR Text Key304371645
Report Number3014704491-2022-00268
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number383019
Device Lot Number1230634
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-