• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US FMS VUE FLUID MANAGEMENT SYSTEM; DISTENSION UNIT, FLUID, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US FMS VUE FLUID MANAGEMENT SYSTEM; DISTENSION UNIT, FLUID, ARTHROSCOPIC Back to Search Results
Model Number 284002
Device Problem Overfill (2404)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/24/2022
Event Type  malfunction  
Event Description
It was reported by the customer in (b)(6) that during an unknown surgery on (b)(6) 2022, it was observed that there was a problem with the pressure sensor on the fms vue fluid management system device.According to the report, the chamber would fill up faster than it would empty.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The device manufacture date is currently unavailable.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the complaint device is not being returned, therefore unavailable for a physical evaluation.Should the device ever be received back in the future, this complaint file will be reopened at that time and an evaluation will be performed and documented.A manufacturing record evaluation was performed for the finished device serial number (b)(6), and no non-conformance was identified.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS VUE FLUID MANAGEMENT SYSTEM
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key14980270
MDR Text Key303579333
Report Number1221934-2022-02069
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705020591
UDI-Public10886705020591
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K130169
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number284002
Device Catalogue Number284002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-