NALU MEDICAL INC NALU NEUROSTIMULATION SYSTEM; SPINAL CORD STIMULATOR Back to Search Results
Model Number 74001
Device Problem Patient-Device Incompatibility (2682)
Patient Problem Inadequate Pain Relief (2388)
Event Date 06/27/2022
Event Type  Injury  
Event Description
Nalu was made aware of a planned revision on (b)(6) 2022.On (b)(6) 2021 the patient reported that when sitting down, the nalu external transmitter modules (etms) will disconnect and start beeping.It was noted that the patient's skin is very loose and the external disc moves easily on the skin when the patient changes positions.The nalu representative attempted multiple troubleshooting strategies.Ultimately the physician and patient decided to revise the implant to a different style that does not require external components.Nalu became aware of this decision on (b)(6) 2022.No other information is available as to status or outcome of the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NALU NEUROSTIMULATION SYSTEM
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
NALU MEDICAL INC
2320 faraday avenue
suite 100
carlsbad CA 92008 2377
Manufacturer (Section G)
NALU MEDICAL, INC
2320 faraday ave
suite 100
carlsbad CA 92008 7241
Manufacturer Contact
terry villarba
2320 faraday avenue
suite 100
carlsbad, CA 92008-7241
7604482360
MDR Report Key14982692
MDR Text Key295674985
Report Number3015425075-2022-00039
Device Sequence Number1
Product Code GZB
UDI-Device Identifier00812537035444
UDI-Public01008125370354441121071310M1706M6
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183579
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number74001
Device Catalogue Number74001
Device Lot NumberM1706 / M6219
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/13/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-