MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ACUITY X4 STRAIGHT; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ACUITY X4 STRAIGHT; IMPLANTABLE LEAD Back to Search Results
Model Number 4671
Device Problems Use of Device Problem (1670); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/24/2022
Event Type  Injury  
Event Description
It was reported that this left ventricular (lv) lead was explanted due to concerns of not being in a good position.A new lead was implanted.No additional adverse patient effects were reported.
 
Event Description
It was reported that this left ventricular (lv) lead was suspected not to be in an optimal position since this patient was not responding as expected to therapy programmed.A chest x ray was performed and confirmed this lead to be in an anterior position.A lead revision was performed and the physician opted to place a new lv lead in a better vein.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUITY X4 STRAIGHT
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14983101
MDR Text Key295671551
Report Number2124215-2022-25118
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526524516
UDI-Public00802526524516
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P010012/S398
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/04/2021
Device Model Number4671
Device Catalogue Number4671
Device Lot Number818229
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/04/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age52 YR
Patient SexMale
-
-